Durag Group GRIMM EDM 280 Environmental Dust Monitor

The GRIMM EDM 280 is an environmental dust monitor designed for real-time PM monitoring in ambient air. This latest-generation optical aerosol spectrometer uses light scattering at single particles to provide precise, continuous measurement of particulate matter under a wide range of environmental conditions. DURAG highlights the EDM 280 for its low detection limit, excellent counting efficiency, and suitability for both fixed monitoring stations and mobile air quality applications.

The EDM 280 is designed to measure TSP, PM10, PM4, PM2.5, PM1, and PMcoarse, while also providing a full particle size distribution across 72 size channels. DURAG states that the instrument is QAL1 certified by TÜV and MCERTS for PM10 and PM2.5 in accordance with EN 12341, EN 16450, EN 15267, and VDI 4202-3, and it is also approved by LCSQA for regulatory air quality monitoring networks. These features make the EDM 280 a strong fit for ambient air monitoring programs where data quality, fast response, and compliance performance are important.

Available in Canada through Westech Industrial, the GRIMM EDM 280 complements Westech’s environmental and analytical monitoring offering for customers who require dependable particulate monitoring in ambient air applications. Westech supports customers across Canada with industrial monitoring solutions and technical product support.

Quantity

PRODUCT FEATURES

REQUEST FOR INFORMATION

Description

  • Real-time ambient air PM monitoring
  • Optical aerosol spectrometer using light scattering at single particles
  • Measures TSP, PM10, PM4, PM2.5, PM1, and PMcoarse
  • Full particle size distribution in 72 size channels
  • QAL1-certified for PM10 and PM2.5
  • Certified for 1-minute measurement intervals
  • Low-maintenance design with no consumables required
  • Optional field verification and calibration kit for on-site quality assurance
  • Compact 19-inch rack design for AQMS cabinets or mobile monitoring laboratories
  • Fast installation and easy removal from rack without dismantling the sampling system.

Documents

  • Will go here

Manufacturer Info

The DURAG Group is a global provider of intelligent solutions for process and environmental monitoring, combustion and safety monitoring, gas analysis, and continuous emissions monitoring systems. With decades of industry experience, DURAG supports safer, more efficient, and more sustainable industrial operations.

Website: https://www.durag.com/

SPECS

Detection principle:
Light scattering at single particles with diode laser; aerodynamically focused detection volume compliant with ISO 21501-1
Measured mass fractions:
TSP, PM10, PM4, PM2.5, PM1, PMcoarse
Certifications:
TÜV and MCERTS QAL1 certified, LCSQA approved
Particle size range:
0.178 μm to 29.4 μm
Size channels:
72
Mass concentration range:
0 to 12,000 μg/m³ for PM10 and 0 to 5,100 μg/m³ for PM2.5
Detection limit:
0.1 μg/m³
Power consumption:
Max. 10 W DC/AC
Sample volume flow:
1.2 l/min
Communication interfaces:
RS-232, USB-B, Ethernet, USB flash drive
Power supply:
100 to 240 VAC + 24 V, 50 to 60 Hz
Ambient site conditions:
-40 to 60 °C
19-inch spectrometer dimensions:
180.5 x 434 x 320 mm

PRODUCT FAQ's

What is the GRIMM EDM 280 used for?
The GRIMM EDM 280 is used for real-time particulate matter monitoring in ambient air, including regulatory air quality monitoring, source apportionment, and mobile monitoring applications.
What particulate fractions can the EDM 280 measure?
The EDM 280 measures TSP, PM10, PM4, PM2.5, PM1, and PMcoarse, and it also provides particle size distribution data across 72 size channels.
Is the D-FW 440 D suitable for compliance-related monitoring?
Yes. DURAG states that the D-FW 440 D is certified to EN 14181, EN 15859, and EN 15267-1/-2, and is also described as US EPA capable, making it suitable for facilities that require recognized dust filter monitoring standards.
Is the EDM 280 suitable for regulatory air quality monitoring?
Yes. DURAG states that the EDM 280 is QAL1-certified by TÜV and MCERTS for PM10 and PM2.5, and approved by LCSQA for use in regulatory air quality monitoring networks.
What makes the EDM 280 different from other PM monitors?
DURAG highlights its 1-minute certified measurement interval, wide particle size range, low detection limit, and aerodynamically focused detection volume that helps avoid border zone errors. It also uses a monochromatic laser to reduce the influence of particle composition on results.
Can Westech help determine if the EDM 280 is the right fit for an ambient air monitoring project in Canada?
Yes. Westech Industrial can help customers review monitoring requirements, installation needs, and application goals to determine whether the GRIMM EDM 280 is a good fit for their ambient air particulate monitoring program.